Fluke Corporation MET/CAL Procedure ============================================================================= INSTRUMENT: Fluke 8010A: (1 year) CAL VER /5700 INSTRUMENT: Fluke 8012A: (1 year) CAL VER /5700 DATE: 09-Feb-95 AUTHOR: Fluke Corporation REVISION: $Revision: 1.6 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 46 NUMBER OF LINES: 179 CONFIGURATION: Fluke 5700A ============================================================================= # # Source: # Fluke 8010A/8012A Digital Multimeter Instrucion Manual # (August 1978, Rev. 2 1/85) # # Compatibility: # MET/CAL 4.0 or later # # Subprocedures: # None # # Required Files: # 57_8010.bmp # 57_8012.bmp # 57CW8012.bmp # 57LO8012.bmp # 57HA8010.bmp # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5700A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # # The 90 day specifications of the 5700A are used in TUR computations. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 MMFC C2 1.002 ASK- P F W 1.003 ASK+ K 1.004 HEAD PRELIMINARY INSTRUCTIONS 1.005 DISP [32] WARNING 1.005 DISP HIGH VOLTAGE is used or exposed during the performance 1.005 DISP of this calibration. DEATH ON CONTACT may result if 1.005 DISP personnel fail to observe safety precautions. 1.006 DISP Before performing verification allow the UUT to 1.006 DISP stabilize at 23degC (+/- 5degC). humidity less than 80% 1.007 DISP Turn on UUT. 1.008 HEAD {DISPLAY TEST} 1.009 JMPT 1.012 Fluke 8012A: (1 year) CAL VER /5700 1.010 PIC 57_8010 1.011 JMP 1.013 1.012 PIC 57CW8012 1.013 5700 20 -18.88V N 2W 1.014 EVAL Are {all segments of the digits illuminated}? 2.001 RNG 2000Z 2.002 EVAL Is the UUT displaying the {overrange indicator} ( 1 )? 3.001 HEAD DISPLAY TEST: {Volts, No input} 3.002 RNG 1000V 3.003 EVAL Is the UUT displaying {000}? 4.001 HEAD DISPLAY TEST: {200mV Range, Short} 4.002 RNG 200mV 4.003 5700 0Z S CW 4.004 EVAL Is the UUT displaying {00.0}? 5.001 HEAD DISPLAY TEST: {2V Range, Short} 5.002 RNG 2V 5.003 5700 0Z S CW 5.004 EVAL Is the UUT displaying {.000}? 6.001 HEAD DISPLAY TEST: {20V Range, Short} 6.002 RNG 20V 6.003 5700 0Z S CW 6.004 EVAL Is the UUT displaying {0.00}? 7.001 HEAD DISPLAY TEST: {200V Range, Short} 7.002 RNG 200V 7.003 5700 0Z S CW 7.004 EVAL Is the UUT displaying {00.0}? 8.001 HEAD DISPLAY TEST: {1000V Range, Short} 8.002 RNG 1000V 8.003 5700 0Z S CW 8.004 EVAL Is the UUT displaying {000}? 9.001 JMPT 9.004 Fluke 8012A: (1 year) CAL VER /5700 9.002 HEAD DISPLAY TEST: {10A/20 MOhm Range, Short} 9.003 JMP 9.005 9.004 HEAD DISPLAY TEST: {20 MOhm Range, Short} 9.005 RNG 20MZ 9.006 5700 0Z S CW 9.007 EVAL Is the UUT displaying {0.00}? 10.001 HEAD {RESISTANCE TEST} 10.002 HEAD RESISTANCE TEST: {200 Ohm Range} 10.003 5700 200 0.0Z +0.2U CW 11.001 HEAD RESISTANCE TEST: {200 Ohm Range} 11.002 5700 200 100.0Z 0.3U CW 12.001 HEAD RESISTANCE TEST: {2 kOhm Range} 12.002 5700 2 1.000kZ 0.003U CW 13.001 HEAD RESISTANCE TEST: {20 kOhm Range} 13.002 5700 20 10.00kZ 0.03U CW 14.001 HEAD RESISTANCE TEST: {200 kOhm Range} 14.002 5700 200 100.0kZ 0.3U 2W 15.001 HEAD RESISTANCE TEST: {2000 kOhm Range} 15.002 5700 2000 1000kZ 6U 2W 16.001 HEAD RESISTANCE TEST: {20 MOhm Range} 16.002 5700 20 10.00MZ 0.06U 2W 17.001 HEAD {CONDUCTANCE TEST} 17.002 HEAD CONDUCTANCE TEST: {2mS Range, Open} 17.003 RNG 2mY 17.004 EVAL Is the UUT displaying {.000}? 18.001 HEAD CONDUCTANCE TEST: {20uS Range, Open} 18.002 RNG 20uY 18.003 EVAL Is the UUT displaying {0.00}? 19.001 HEAD CONDUCTANCE TEST: {200nS Range, Open} 19.002 RNG 200nY 19.003 MATH MEM1 = 0.0 19.004 MEMI Enter UUT reading in nanosiemens: 19.005 MEME 19.006 MEMC 200 nY +0.3U 20.001 JMPT 22.002 Fluke 8010A: (1 year) CAL VER /5700 20.002 HEAD RESISTANCE TEST: {LO RANGE Ohm} 20.003 DISP Select LO RANGE OHMS on the UUT. 20.004 PIC 57LO8012 20.005 HEAD RESISTANCE TEST: {2 Ohm Range} 20.006 RNG 2Z 20.007 5700 0Z S RW 20.008 DISP Adjust the ZERO control for a UUT reading of all zeros. 20.009 5700 2 1.9Z 0.021U RW 21.001 HEAD RESISTANCE TEST: {20 Ohm Range} 21.002 5700 20 19Z 0.12U RW 22.001 PIC 57_8012 22.002 HEAD {DC VOLTAGE TEST} 22.003 HEAD DC VOLTAGE TEST: {200mV Range} 22.004 5700 200 190.0mV 0.3U 2W 23.001 HEAD DC VOLTAGE TEST: {2V Range} 23.002 5700 2 1.900V 0.003U 2W 24.001 HEAD DC VOLTAGE TEST: {20V Range} 24.002 5700 20 19.00V 0.03U 2W 25.001 5700 20 -19.00V 0.03U 2W 26.001 HEAD DC VOLTAGE TEST: {200V Range} 26.002 5700 200 190.0V 0.3U 2W 27.001 HEAD DC VOLTAGE TEST: {1000V Range} 27.002 5700 1000 1000V 2U 2W 28.001 HEAD {AC VOLTAGE TEST} 28.002 HEAD AC VOLTAGE TEST: {200mV Range} 28.003 5700 200 190.0mV 1.2U 100H 2W 29.001 5700 200 10.0mV 0.2U 100H O 2W 30.001 HEAD AC VOLTAGE TEST: {2V Range} 30.002 5700 2 1.900V 0.012U 100H 2W 31.001 HEAD AC VOLTAGE TEST: {20V Range} 31.002 5700 20 19.00V 0.12U 100H 2W 32.001 5700 20 1.00V 0.03U 10kH 2W 33.001 5700 20 19.00V 0.21U 10kH 2W 34.001 5700 20 19.00V 0.97U 50kH 2W 35.001 HEAD AC VOLTAGE TEST: {200V Range} 35.002 5700 200 190.0V 1.2U 100H 2W 36.001 5700 200 110.0V 1.3U 10kH 2W 37.001 HEAD AC VOLTAGE TEST: {750V Range} 37.002 5700 750 750V 6U 100H 2W 38.001 5700 750 250V 3U 1kH 2W 39.001 HEAD {DC CURRENT TEST} 39.002 HEAD DC CURRENT TEST: {200uA Range} 39.003 5700 200 190.0uA 0.7U 2W 40.001 HEAD DC CURRENT TEST: {2mA Range} 40.002 5700 2 1.900mA 0.007U 2W 41.001 HEAD DC CURRENT TEST: {20mA Range} 41.002 5700 20 19.00mA 0.07U 2W 42.001 HEAD DC CURRENT TEST: {200mA Range} 42.002 5700 200 190.0mA 0.7U 2W 43.001 HEAD DC CURRENT TEST: {2000mA Range} 43.002 5700 2000 1900mA 7U 2W 44.001 HEAD {AC CURRENT TEST} 44.002 HEAD AC CURRENT TEST: {200uA Range} 44.003 5700 200 190.0uA 2.1U 1kH 2W 45.001 JMPT 46.001 Fluke 8012A: (1 year) CAL VER /5700 45.002 DISP Remove all connections. 45.003 HEAD {10A CURRENT TEST} 45.004 PIC 57HA8010 45.005 5700 10 1.90A 0.02U 2W 46.001 END